Case study on the orientation of phaco hand pieces during steam sterilization processes
J P C M van Doornmalen Gomez Hoyos1, R A C van Wezel2, H W J M van Doornmalen3
1Eindhoven University of Technology, Department of Applied Physics, Eindhoven, The Netherlands; 3M Deutschland GmbH, Infection Prevention Division, Europe, Middle East & Africa, Neuss, Germany.
Background:
Steam sterilization is an essential part of infection prevention. The literature shows that sterilization of medical instruments containing channels is not trivial. Phaco hand pieces have a simple configuration: a device contains a channel with a constant radius. No literature was found indicating whether the sterilization conditions on the inner surface of a phaco hand piece are influenced by the orientation of the hand piece.
Aim:
To determine whether the orientation of a phaco hand piece influences the results of a sterilization process of this device.
Methods:
A qualitative case study, including experiments, is performed with a protocolled combination of steam sterilizer, process, phaco hand piece, orientation of the phaco hand piece, and wrapping.
Findings:
In this specific case, the orientation of the hand piece influenced the result of the steam sterilization process; in vertically (upright) oriented phaco hand pieces with free water drainage, sterilization conditions are reproducibly established. In the same process, in horizontally oriented or vertically oriented hand pieces without free drainage, these conditions are not established in a reproducible way.
Conclusion:
In the investigated combination of sterilizer, process, load, loading pattern and wrapping, phaco hand pieces have to be oriented vertically (upright) with free water drainage to obtain steam sterilization conditions on the inner surface. It is likely that instruments with comparable configuration and dimensions will yield comparable results. It is therefore recommended that this issue is considered during the development of medical instruments and during performance qualifications of such instruments.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Hand hygiene
Hand washing...
Handwashing III: During the Procedure and Post-Procedure Steps
Handwashing I: Introduction and Types of Equipment
Hand wash basins in clinical areas should have faucets that can be turned on and off without using the hands; that is, they should be non-touch or lever-operated....