How to Appropriately Calculate Effective Dose for CT Using Either Size-Specific Dose Estimates or Dose-Length Product
Samuel L Brady1, Amy E Mirro, Bria M Moore
11 Department of Radiological Sciences, St Jude Children's Research Hospital, 262 Danny Thomas Pl, Memphis TN, 38105.
Objective:
The purpose of this study is to show how to calculate effective dose in CT using size-specific dose estimates and to correct the current method using dose-length product (DLP).
Materials And Methods:
Data were analyzed from 352 chest and 241 abdominopelvic CT images. Size-specific dose estimate was used as a surrogate for organ dose in the chest and abdominopelvic regions. Organ doses were averaged by patient weight-based populations and were used to calculate effective dose by the International Commission on Radiological Protection (ICRP) report 103 method using tissue-weighting factors (EICRP). In addition, effective dose was calculated using population-averaged CT examination DLP for the chest and abdominopelvic region using published k-coefficients (EDLP = k × DLP).
Results:
EDLP differed from EICRP by an average of 21% (1.4 vs 1.1) in the chest and 42% (2.4 vs 3.4) in the abdominopelvic region. The differences occurred because the published kcoefficients did not account for pitch factor other than unity, were derived using a 32-cm diameter CT dose index (CTDI) phantom for CT examinations of the pediatric body, and used ICRP 60 tissue-weighting factors. Once it was corrected for pitch factor, the appropriate size of CTDI phantom, and ICRP 103 tissue-weighting factors, EDLP improved in agreement with EICRP to better than 7% (1.4 vs 1.3) and 4% (2.4 vs 2.5) for chest and abdominopelvic regions, respectively.
Conclusion:
Current use of DLP to calculate effective dose was shown to be deficient because of the outdated means by which the k-coefficients were derived. This study shows a means to calculate EICRP using patient size-specific dose estimate and how to appropriately correct EDLP.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Dosing: Infants and Children
Biological Effects of Radiation
Dose Response Curve: Conventional Versus Nonmonotonic
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant