Comparison of the QMS Analyzer With HPLC-UV for the Quantification of Lamotrigine Concentrations in Human Plasma
Sara Baldelli1, Simone Castoldi, Nitin Charbe
1*Unit of Clinical Pharmacology, Luigi Sacco University Hospital, Milan; †Clinical Pharmacology Unit, Consiglio Nazionale delle Ricerche Institute of Neuroscience, Department of Biomedical and Clinical Sciences, Luigi Sacco University Hospital, Università degli Studi di Milano, Milan; and ‡Eugenio Medea Scientific Institute, Bosisio Parini, Italy.
Background:
Recently, a turbidimetric immunoassay method has been developed for use in the form of a QMS lamotrigine (LTG) commercial immunoassay. This study was designed to evaluate the performance of this immunoassay using a validated high-performance liquid chromatography-ultraviolet (HPLC-UV) method as the reference.
Methods:
The performance of QMS was initially tested using drug-free plasma spiked with different amounts of LTG and, subsequently, by analyzing 61 trough plasma samples from epileptic patients given the drug as part of their maintenance antiepileptic therapies.
Results:
The correlation between LTG concentrations measured by QMS and HPLC was good, with a Pearson coefficient of 0.968 (P < 0.0001). The Bland-Altman approach showed that LTG concentrations measured with QMS exceeded HPLC on an average by 15.6% (limits of agreement, -18% to +63%), with a concentration-dependent performance (mean percent bias, 49.5 ± 8.2% and 0.6 ± 12.7% for concentrations less than 2 mg/L and greater than 14.9 mg/L, respectively).
Conclusions:
The QMS provided acceptable analytical performance across a wide concentration range for routine LTG measurements, being at least comparable with the other commercial immunoassays. It could be, therefore, considered as a viable alternative to HPLC methods for routine LTG monitoring in the clinical practice, although its suitability for accurate analysis of samples with low concentration is limited.
More Related Videos
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024
07:34Large Scale Non-targeted Metabolomic Profiling of Serum by Ultra Performance Liquid Chromatography-Mass Spectrometry UPLC-MS
Published on: March 14, 2013
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Mass Analyzers: Common Types
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
High-Resolution Mass Spectrometry (HRMS)