Related Experiment Video
Updated: Apr 16, 2026

A Novel Approach for the Administration of Medications and Fluids in Emergency Scenarios and Settings
Published on: November 9, 2016
Maji: a new tool to prevent overhydration of children receiving intravenous fluid therapy in low-resource settings
Kamal Shah1, Erica Skerrett1, Matthew Nojoomi1
1Department of Bioengineering, Rice University, Houston, Texas; Rice 360°, Institute for Global Health Technologies, Rice University, Houston, Texas; 3rd Stone Design, San Rafael, California; Department of Paediatrics, Queen Elizabeth Central Hospital, Blantyre, Malawi.
Insights
A new mechanical device accurately regulates fluid volume during intravenous therapy, preventing overhydration in children. This crucial tool enhances patient safety in resource-limited settings lacking advanced infusion equipment.
Area of Science:
- Biomedical Engineering
- Pediatric Nursing
- Medical Device Development
Background:
- Intravenous (IV) fluid therapy is critical for pediatric care, especially in sub-Saharan Africa.
- Overhydration is a significant risk in children receiving IV fluids due to the lack of precise volume control devices like burettes or electronic infusion pumps.
- Clinicians require accessible tools to manage IV fluid administration accurately in resource-limited pediatric settings.
Purpose of the Study:
- To design and evaluate a novel mechanical device for regulating fluid volume during intravenous drip therapy.
- To assess the accuracy and usability of this device for preventing overhydration in pediatric patients.
- To provide a solution for accurate IV fluid management in settings lacking advanced infusion technology.
Main Methods:
- A mechanical fluid regulation system was designed for compatibility with standard IV bags.
- The device limits dispensed volume to a preset, adjustable amount (in 50 mL increments).
- Laboratory testing evaluated accuracy across various flow rates, initial bag volumes, and target volumes.
Main Results:
- The device demonstrated laboratory accuracy within 12.0 mL of the target volume across tested conditions.
- Ease of use was rated as "excellent," with a mean System Usability Scale score of 84.4/100.
- The device effectively limits dispensed IV fluid volume.
Conclusions:
- The developed mechanical device accurately and reliably regulates IV fluid volume.
- Its excellent usability and accuracy make it a valuable tool for preventing pediatric overhydration in resource-limited environments.
- This innovation has the potential to reduce morbidity and mortality associated with IV fluid therapy complications.
Abstract:
We designed and evaluated the accuracy and usability of a device to regulate the volume of fluid dispensed during intravenous drip therapy. The mechanical system was developed in response to a pressing need articulated by clinicians in pediatric wards throughout sub-Saharan Africa, who require a tool to prevent overhydration in children receiving intravenous fluid in settings that lack burettes or electronic infusion pumps. The device is compatible with most intravenous bags and limits the volume dispensed to a preset amount that can be adjusted in 50 mL increments. Laboratory accuracy over a range of clinically-relevant flow rates, initial bag volumes, and target volumes was within 12.0 mL of the target volume. The ease of use is "excellent," with a mean system usability score of 84.4 out of 100. Use of the device limits the volume of fluid dispensed during intravenous therapy and could potentially reduce the morbidity and mortality associated with overhydration in children receiving intravenous therapy.
Related Concept Videos
Drug Dosing: Infants and Children
Hemodialysis III: Nursing Management
IV Infusion to Oral Dosing: Conversion Methods
Acute Kidney Injury VI: Nursing Management
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Regulation of Water Intake