Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Size-specific Dose Estimates for Chest, Abdominal, and Pelvic CT: Effect of Intrapatient Variability in
Shuai Leng1, Maria Shiung1, Xinhui Duan1
1From the Department of Radiology, Mayo Clinic, 200 First St SW, Rochester, MN 55905.
Purpose:
To develop software to automatically calculate size-specific dose estimates (SSDEs) and to assess the effect of variations in water-equivalent diameter (Dw) along the z-axis on SSDE for computed tomographic (CT) examinations of the torso.
Materials And Methods:
In this institutional review board-approved, HIPAA-compliant, retrospective study, a software program was used to calculate Dw at each image position in 102 consecutive CT examinations of the combined chest, abdomen, and pelvis. SSDE was calculated by multiplying the size-dependent conversion factor and volume CT dose index (CTDIvol) at each image position. The variations in Dw along the z-axis were determined for six hypothetical scanning ranges: chest alone; abdomen alone; pelvis alone; chest and abdomen; abdomen and pelvis; and chest, abdomen, and pelvis. Mean SSDE was calculated in two ways: (a) from the SSDE at each position and (b) from the mean CTDIvol over each scan range and the conversion factor corresponding to Dw at the middle of the scan range. Linear regression analysis was performed to determine the correlation between SSDE values calculated in these two ways.
Results:
Across patients, for scan ranges 1-6, the mean of the difference between maximal and minimal Dw within a given patient was 5.2, 4.9, 2.5, 6.0, 5.6, and 6.5 cm, respectively. The mean SSDE values calculated by using the two methods were in close agreement, with root mean square differences of 0.9, 0.5, 0.5, 1.4, 1.0, and 1.1 mGy or 6%, 3%, 2%, 9%, 4%, and 6%, for the scan ranges of chest; abdomen; pelvis; chest and abdomen; abdomen and pelvis; and chest, abdomen, and pelvis, respectively.
Conclusion:
Using the mean CTDIvol from the whole scan range and Dw from the image at the center of the scan range provided an easily obtained estimate of SSDE for the whole scan range that agreed well with values from an image-by-image approach, with a root mean square difference less than 1.4 mGy (9%).
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
09:00Spatial Measurements of Perfusion, Interstitial Fluid Pressure and Liposomes Accumulation in Solid Tumors
Published on: August 18, 2016
Related Concept Videos
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...