Estimation of radioactivity in single-photon emission computed tomography for sentinel lymph node biopsy in a torso
Hideaki Kitamura1, Yuzuru Kono, Takeshi Murano
1aDiagnostics Radiology Division, National Cancer Center Hospital bGraduate School of Human Health Sciences, Tokyo Metropolitan University, Tokyo, Japan.
Objective:
The number of lymph nodes to be removed is determined from residual counts. Advance estimation of residual radioactivity in lymphatic nodes before a biopsy is useful for reducing surgical operation time. The purpose of this study was to estimate the total radioactivity of a small hotspot in single-photon emission computed tomography (SPECT) of a torso phantom.
Methods:
A cross-calibration study was performed to convert counts in SPECT images to radioactivity. A simulation study was performed to estimate the size of the volume of interest (VOI) covering a hotspot corrupted with full-width at half-maximum between 8 and 16 mm. The estimation of total radioactivity was validated in a torso phantom study using small sources.
Results:
True radioactivity was approximately equal to integrated values of hotspots using the VOI with a diameter of 40 mm in our simulation study. The difference was less than 18% in cases of more than 9.4 kBq.
Conclusion:
The total radioactivity in small sources simulating a typical sentinel node was estimated from SPECT images using a VOI of 40 mm in a torso phantom study. Because the difference from actual values was less than 10% on average when radioactivities were more than 9.4 kBq, the total radioactivity of a lymph node can be estimated in a clinical examination.