Validation of noninvasive hemoglobin measurement by pulse co-oximeter in newborn infants
C Nicholas1, R George2, S Sardesai3
11] Division of Neonatal Medicine, Department of Pediatrics, LAC+USC Medical Center, University of Southern California, Los Angeles, CA, USA [2] Division of Neonatal Medicine, Department of Pediatrics, Children's Hospital Los Angeles, Keck School of Medicine, University of Southern California, Los Angeles, CA, USA.
Objective:
To describe the accuracy of noninvasive hemoglobin (Hb) obtained with pulse co-oximeter (SpHb) compared with total Hb (tHb) from laboratory co-oximeter in neonates.
Study Design:
Neonates with birth weight (BW) <3000 g admitted to LAC+USC Medical Center neonatal intensive care unit were included. SpHb was recorded using Masimo Radical-7 and compared with tHb. A total of three data sets were obtained for each patient. Regression analysis and Bland-Altman analysis were performed.
Result:
Sixty-one patients (mean±s.d., BW 1177±610 g and gestational age 28.7±3.9 weeks) were enrolled. The mean tHb value was 13.9±2.0 g dl(-1) and the mean SpHb was 14.0±2.0 g dl(-1). There was a moderately positive correlation between SpHb and tHb (r=0.66, P<0.001) with a bias and precision of -0.09±1.67 g dl(-1). Data from a subgroup of infants with gestational age ⩽32 weeks (52/61 patients) were analyzed, and the correlation coefficient was moderately positive (r=0.69, P<0.001) with a bias and precision of -0.23±1.60 g dl(-1).
Conclusion:
Our results suggest that noninvasive SpHb may be considered as an adjunct to invasive tHb measurements in newborn infants <3000 g especially in preterm infants ⩽32 weeks of gestation.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...