Optimal pacing sites for cardiac resynchronization therapy for patients with a systemic right ventricle with or
Aya Miyazaki1, Heima Sakaguchi2, Koji Kagisaki3
1Department of Pediatric Cardiology, National Cerebral and Cardiovascular Center, 5-7-1 Fujishirodai, Suita, Osaka 565-8565, Japan ayamiya@ncvc.go.jp.
Aims:
This study aims to assess the impact of pacing sites on the effectiveness of cardiac resynchronization therapy (CRT) in systemic right ventricle (sRV) patients with/without a rudimentary left ventricle (rLV).
Methods And Results:
We evaluated 13 procedures in 11 sRV patients with a wide QRS (>150 ms). Based on the digitalization results of ventriculography, long-axis dyssynchrony (LD) was defined as extremely delayed right ventricular (RV) outflow tract movement: ≥100 ms delay from the RV apical contraction, and short-axis dyssynchrony (SD) was defined as a paradoxical contraction between the rLV and sRV caused by a conduction delay between the two ventricles. During the follow-up period (2.1 ± 1.9 years), the response rates were 71% (5/7) and 33% (2/6) in the sRV patients with and without an rLV, respectively (P = ns). Following the CRT, the QRS duration remained similar between the responders and nonresponders. Among five responders with an rLV, the leads were placed in the longitudinal RV direction in two with LD, longitudinal RV direction with fusion of the intrinsic QRS in two with LD + SD, and laterally on opposite sides of both ventricles in one with SD. Among two responders without an rLV, the leads were placed in the longitudinal RV direction in those two with LD.
Conclusions:
In sRV patients with LD with/without an rLV, the leads should be placed at furthest sites in the longitudinal RV direction. In patients with an rLV and SD, the leads should be placed laterally on opposite sides of both ventricles.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Cardiac Catheterization II: Right Heart Catheterization
Cardiomyopathy V: Interprofessional Care
Cardiomyopathy II: Dilated Cardiomyopathy
Cardiomyopathy III: Hypertrophic Cardiomyopathy
Conduction System of the Heart
This system relies on the unique properties of nodal and Purkinje cells:...
Conduction System of the Heart
The pacemaker cells are located in two primary nodes: the sinoatrial (SA) node and the atrioventricular (AV) node. The SA node pacemaker cells can autonomously depolarize, triggering an action potential that leads to the...