Evaluation of the Microstat™ sublingual PCO2 monitor in ambulatory patients
Benjamin D Fox1,2, Dominique Joyal3, Robert D Schlesinger4
1Center for Pulmonary Vascular Disease, Divisions of Cardiology, Lady Davis Institute for Medical Research, Jewish General Hospital, McGill University, Room E-222, 3755 Cote Ste Catherine, Montreal, QC, H3T 1E2, Canada. benfox@post.tau.ac.il.
Abstract:
Physicians often need to measure arterial PCO2 in clinical practice. Arterial blood gas sampling is typically available only in hospitals and may be unpleasant for patients. Minimally invasive techniques for measuring PCO2 offer the potential for overcoming these limitations. The MicroStat monitor non-invasively measures PCO2 in the sublingual tissues, which should track arterial PCO2 in hemodynamically stable patients. This was a prospective observational study. Patients undergoing routine cardiac catheterization were recruited. Following arterial cannulation, two sequential sublingual PCO2 measurements were taken and a contemporaneous arterial sample was sent for blood gas analysis. For each subject we calculated the mean sublingual-arterial CO2 gradient and the test-retest sublingual PCO2 difference. Twenty-five patients were studied. Mean sublingual-arterial PCO2 gradient was +6.8 mmHg (95 % limits of agreement -3.0 to 16.6 mmHg). Test-retest difference was 3.4 mmHg (95 % limits of agreement -1.1 to 7.9 mmHg), p = 0.11 (Wilcoxon test), repeatability was 11 mmHg. The MicroStat sublingual PCO2 monitor over-estimates arterial PCO2 with wide limits of agreement. Test-retest repeatability was poor. Use of sublingual PCO2 monitoring with the MicroStat monitor cannot currently replace blood gas sampling.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Diagnosing Acidosis and Alkalosis
First, the pH level is assessed to determine whether the blood pH is normal (7.35–7.45), low (acidosis), or high (alkalosis).
Next, the PCO2 and...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: