Comparative evaluation of 2.0 mm locking plate system vs 2.0 mm non-locking plate system for mandibular angle
1Departement of Stomatology, The Second Hospital of Lanzhou University, 82 Cuiyingmwen, Chengguan District, Lanzhou City, Gansu Province, China. pavanpatil2000@yahoo.co.uk.
Objective:
This prospective randomized study evaluated the efficacy of a 2.0 mm locking plate/screw system compared with a 2.0-mm nonlocking plate/screw system in fixation of 60 isolated non-comminuted mandibular angle fractures.
Patients And Methods:
Sixty patients were randomly assigned to receive a 2.0 mm locking plate (group A, n = 30) or 2.0 mm non-locking plate (group B, n = 30). All patients were followed up to 6 months postoperatively and evaluated for complications, occlusal stability and overall results of fixation.
Results:
Five complications occurred in the locking group and fourteen in the non-locking group with complication rates equalling 17% and 47% respectively. When comparing the overall results according to plates used, the χ² test showed a statistically significant difference between the locking and non-locking plates (p < 0.01). Fewer patients required IMF in group A.
Conclusions:
Mandibular angle fractures treated with 2.0 mm locking plates show greater stability and were associated with fewer complications than with 2.0 mm non-locking plates.