Related Experiment Video
Updated: Apr 16, 2026

A Novel Non-invasive Method for the Detection of Elevated Intra-compartmental Pressures of the Leg
Published on: May 31, 2019
Interobserver Variability in the Measurement of Lower Leg Compartment Pressures
Thomas M Large1, Julie Agel, Daniel J Holtzman
1*Orthopaedic Trauma Services, Mission Hospital, Asheville, NC; †Department of Orthopaedic Surgery and Sports Medicine, Harborview Medical Center, Seattle, WA; and ‡Rothman Institute at Thomas Jefferson University, Philadelphia, PA.
Objectives:
To determine whether interobserver technical variations and errors in the measurement of compartment pressures may affect measurement accuracy.
Methods:
Four above-knee cadaveric specimens were used to create a consistent model of lower leg compartment syndrome. Thirty-eight physicians examined the limbs and measured 4 compartment pressures using the Intra-Compartmental Pressure Monitor (Stryker Orthopaedics). They were observed for correct assembly and use of the monitor. Measurements obtained were compared with known pressures.
Results:
Of the total number of compartment measurements, 31% were made using the correct technique, 39% were made with lesser errors in technique, and 30% were made with catastrophic errors. Only 60% of measurements made with the correct technique were within 5 mm Hg of the standard pressure. Accuracy dropped to 42% for measurements taken with small errors in technique and 22% when a catastrophic error was committed.
Conclusions:
Variations in use of a commercially available pressure monitor exist, and errors are common. Proper use improved accuracy, but even with proper technique, 40% of the measurements were >5 mm Hg from the actual pressure. Based on our data, measurement accuracy with this device should be questioned and viewed within a range. Regular review and education of technique is strongly recommended.
More Related Videos
Related Concept Videos
Assessing Blood pressure in the Leg
Preparation:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure