Related Experiment Video
Updated: Apr 16, 2026

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Cyclic fatigue resistance of 3 different nickel-titanium reciprocating instruments in artificial canals
Oscar Higuera1, Gianluca Plotino2, Luigi Tocci2
1Department of Endodontics, Universidad Autonoma de Baja California, Tijuana, Mexico.
Introduction:
The purpose of this study was to evaluate the cyclic fatigue resistance of 3 different nickel-titanium reciprocating instruments.
Methods:
A total of 45 nickel-titanium instruments were tested and divided into 3 experimental groups (n = 15): group 1, WaveOne Primary instruments; group 2, Reciproc R25 instruments; and group 3, Twisted File (TF) Adaptive M-L1 instruments. The instruments were then subjected to cyclic fatigue test on a static model consisting of a metal block with a simulated canal with 60° angle of curvature and a 5-mm radius of curvature. WaveOne Primary, Reciproc R25, and TF Adaptive instruments were activated by using their proprietary movements, WaveOne ALL, Reciproc ALL, and TF Adaptive, respectively. All instruments were activated until fracture occurred, and the time to fracture was recorded visually for each file with a 1/100-second chronometer. Mean number of cycles to failure and standard deviations were calculated for each group, and data were statistically analyzed (P < .05). Instruments were also observed through scanning electron microscopy to evaluate type of fracture.
Results:
Cyclic fatigue resistance of Reciproc R25 and TF Adaptive M-L1 was significantly higher than that of WaveOne Primary (P = .009 and P = .002, respectively). The results showed no statistically significant difference between TF Adaptive M-L1 and Reciproc R25 (P = .686). Analysis of the fractured portion under scanning electron microscopy indicated that all instruments showed morphologic characteristics of ductile fracture that were due to accumulation of metal fatigue.
Conclusions:
No statistically significant differences were found between the instruments tested except for WaveOne Primary, which showed the lowest resistance to cyclic fatigue.
More Related Videos
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
06:48Rapid Formation and Testing of Self-expanding NiTi Frames with a Small Form Factor Suitable for Minimally Invasive Implants
Published on: March 7, 2025