Related Experiment Video
Updated: Apr 16, 2026

Cooling or Warming the Esophagus to Reduce Esophageal Injury During Left Atrial Ablation in the Treatment of Atrial Fibrillation
Published on: March 15, 2020
Comparison between single- and multi-sensor oesophageal temperature probes during atrial fibrillation ablation:
Cory M Tschabrunn1, Joshua Silverstein1, Tyler Berzin2
1Harvard-Thorndike Electrophysiology Institute, Cardiovascular Division, Beth Israel Deaconess Medical Center, Harvard Medical School, 185 Pilgrim Road, Baker 4, Boston, MA 02215, USA.
Aims:
Oesophageal temperature monitoring with single-sensor probe (SSP) has a variable ability to predict oesophageal ulceration as a consequence of pulmonary vein isolation (PVI). Multi-sensor self-expandable probes (MSPs) may offer improved thermal monitoring. The objective of this study was to compare the thermodynamic characteristics of both probes during PVI.
Methods And Results:
This prospective study enrolled 20 patients undergoing index PVI. Ten patients (group A) underwent dual monitoring with SSP and MSP and 10 control patients (group B) were monitored with SSP alone. Time to initial rise (>0.2°C), time to 1.0°C rise, peak temperature, and decay were recorded with each posterior wall lesion (20 W, 198 applications). The operator was blinded to the MSP temperature data and ablation was only interrupted when SSP temperature increased by ≥2°C. All patients underwent endoscopy within 24 h. Initial temperature increase was detected earlier with MSP (13.4 ± 7.5 vs. 30.5 ± 15.4 s; P < 0.001); led to shorter time to 1.0°C rise (18.5 ± 10.1 vs. 32.1 ± 12.0 s; P < 0.001); and higher change in peak temperature (1.6 ± 2.0 vs. 0.60 ± 0.53°C; P < 0.001). Decay time was similar between the probes (146.1 ± 35.3 vs. 150.4 ± 48.4 s; P = 0.89). The incidence of oesophageal ulceration was similar between the Groups A and B (5 and 4, respectively). Multi-sensor self-expandable probe provided greater sensitivity (100 vs. 60%) and similar specificity (60%) for detection of oesophageal ulceration. Five swine underwent oesophageal mapping before and after MSP placement without alteration in size or position.
Conclusion:
Multi-sensor probes provide a superior thermodynamic profile. Its clinical value in reducing oesophageal injury requires further evaluation.
More Related Videos
05:43Author Spotlight: Improving Lesion Contiguity in Pulmonary Vein Isolation via Proactive Esophageal Cooling
Published on: April 19, 2024
04:58Reduced Procedure Time and Variability with Active Esophageal Cooling During Radiofrequency Ablation for Atrial Fibrillation
Published on: August 25, 2022
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Axilla
Step 1: Perform hand hygiene and put on clean gloves to maintain infection control and prevent cross-contamination.
Step 2: Prepare the patient by explaining the procedure to ensure understanding and cooperation. Ensure privacy, expose the axilla, and inform the patient that minimal movement is crucial for an accurate reading.
Step 3: Adjust the patient’s clothing to expose only the axilla. It minimizes...