Factors limiting vocal-tract length discrimination in cochlear implant simulations
Etienne Gaudrain1, Deniz Başkent1
1Department of Otorhinolaryngology/Head and Neck Surgery, University Medical Center Groningen, University of Groningen, Groningen, Netherlands.
Abstract:
Perception of voice characteristics allows normal hearing listeners to identify the gender of a speaker, and to better segregate speakers from each other in cocktail party situations. This benefit is largely driven by the perception of two vocal characteristics of the speaker: The fundamental frequency (F0) and the vocal-tract length (VTL). Previous studies have suggested that cochlear implant (CI) users have difficulties in perceiving these cues. The aim of the present study was to investigate possible causes for limited sensitivity to VTL differences in CI users. Different acoustic simulations of CI stimulation were implemented to characterize the role of spectral resolution on VTL, both in terms of number of channels and amount of channel interaction. The results indicate that with 12 channels, channel interaction caused by current spread is likely to prevent CI users from perceiving VTL differences typically found between male and female speakers.