Related Experiment Video
Updated: Apr 16, 2026

Methods for Quantitative Detection of Antibody-induced Complement Activation on Red Blood Cells
Published on: January 29, 2014
Antibodies reactive to cleaved sites in complement proteins enable highly specific measurement of soluble markers of
Anna M Blom1, Anders Österborg2, Tom E Mollnes3
1Department of Laboratory Medicine Malmö, Section of Medical Protein Chemistry, Lund University, Inga Marie Nilssons Street 53, 20502 Malmö, Sweden.
Abstract:
An emerging number of diseases and therapeutic approaches with defined involvement of the complement system justify a need for specific markers reflecting activation of particular effector arms of the complement cascade. Measurement of such soluble markers in circulation is a challenge since the specificity of antibodies must be limited to activated complement fragments but not predominant and ubiquitous parental molecules. Existing assays for the measurement of soluble, activated complement proteins are based on the detection of conformational neoepitopes. We tested an alternative approach based on detection of short linear neoepitopes exposed at the cleavage sites after activation of the actual complement component. Obtained antibodies reactive to C4d and C5b fragments enabled us to set up highly specific sandwich ELISAs, which ensured trustful measurements without false positive readouts characteristic for some of the widely used assays.
Related Concept Videos
Complement System
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Antibody Actions
Neutralization
Antibodies can bind to pathogens, preventing them from infecting host cells. This process...