Background:
Intravenous iron is commonly administered to anaemic patients to treat iron deficiency, but due to its ferric colouration, it may interfere with the spectrophotometric assessment of haemoglobin concentrations. This paper investigates the potential interference of three clinically used intravenous iron preparations on the measurement of haemoglobin.
Methods:
Haemoglobin concentration was measured for neat and Hartmann's solution-diluted iron polymaltose, carboxymaltose and sucrose solutions using bedside (Radiometer HemoCue®), point-of-care (Radiometer ABL800 Flex) and laboratory (Abbott CellDyne Sapphire™) devices. Haemoglobin concentration was then assessed with the same devices utilizing anaemic whole blood with the iron solutions added.
Results:
Neat iron preparations registered clinically significant haemoglobin concentrations on bedside and laboratory measurements. When intravenous iron preparations were diluted to clinical concentrations, their effect on haemoglobin measurements, either in isolation or mixed with anaemic blood, was negligible.
Conclusion:
Although neat preparations of intravenous iron do interfere with spectrophotometric analysis of haemoglobin, concentrations likely to be seen post iron infusion do not significantly interfere with haemoglobin measurement.
Related Concept Videos
Measurement of Bioavailability: Pharmacodynamic Methods
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Measurement of Bioavailability: Pharmacokinetic Methods
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Effect of Hepatic Disease on Pharmacokinetics: Dose Adjustments Due to Hepatic Impairment
IV Infusion to Oral Dosing: Conversion Methods