Bipolar tined polyurethane ventricular lead: a four-year experience.
Pacing and Clinical Electrophysiology : PACE
|March 1, 1985
Summary
Polyurethane pacemaker leads, particularly bipolar tined ventricular models, show varying failure rates based on manufacturing date. Early leads had an 8.8% failure rate, while later ones had 3.9%, highlighting the need for ongoing lead integrity evaluation.
Area of Science:
- Biomedical Engineering
- Cardiovascular Devices
- Materials Science
Background:
- Polyurethane insulation is increasingly used in permanent pacemaker leads.
- Concerns exist regarding the long-term durability and integrity of these polyurethane-insulated leads.
- Specific attention is drawn to bipolar tined polyurethane ventricular leads due to potential failure risks.
Purpose of the Study:
- To assess the 4-year performance and failure rates of specific bipolar tined polyurethane ventricular pacemaker leads.
- To compare lead integrity between two manufacturing groups: those produced before and after a specified date.
- To evaluate the need for improved lead design, failure rate determination, and registry mechanisms.
Main Methods:
- Retrospective analysis of a 4-year patient experience with polyurethane-insulated pacemaker leads.
- Categorization of patients into two groups based on lead manufacturing date (pre- and post-specified date).
- Calculation of lead failure rates and average time to failure for each group; actuarial analysis for survivorship.
Main Results:
- The failure rate was 8.8% in the earlier manufactured group (at increased risk) versus 3.9% in the later manufactured group (not at increased risk).
- Lead failures in the first group occurred at an average of 17.5 months.
- Actuarial analysis indicated potentially non-significant differences in lead failure-free survivorship between the two groups, pending longer follow-up for the second group.
Conclusions:
- The study underscores the importance of monitoring polyurethane pacemaker lead integrity and failure rates.
- Identifying optimal lead designs and configurations is crucial for improving long-term device performance.
- Establishing a lead registry is recommended for evaluating the ongoing integrity of various pacemaker leads.