Comparative evaluation of 2.3 mm locking plate system vs conventional 2.0 mm non locking plate system for mandibular
1Department of Stomatology, Central Hospital of Xuzhou City, Affiliated to Southeast University, Xuzhou Stomatology Clinical School of Weifang Medical University, Xuzhou Jiangsu Province, China. mrocket@126.com.
Objective:
This retrospective study evaluated the efficacy of a 2.3 mm locking plate/screw system compared with a 2.0-mm non-locking plate/screw system in fixation of isolated non comminuted mandibular condyle fractures.
Patients And Methods:
Surgical records of 101 patients who received either a 2.3 mm locking plate (group A, n = 51) or 2.0 mm non locking plate (group B, n = 50) were analyzed. All patients were followed up to a minimum of 6 months postoperatively and evaluated for hardware related complications, occlusal stability, need for and duration of MMF and mandibular functional results.
Results:
Four complications occurred in the locking group and eighteen in the non locking group with complication rates equalling 8% and 36% respectively. When comparing the overall results according to plates used, the χ2 test showed a statistically significant difference between the locking and non locking plates (p < 0.001). Fewer patients required postoperative MMF in group A.
Conclusions:
Mandibular condyle fractures treated with a 2.3 mm locking plate exhibited stable osteosynthesis, were associated with minimal complications and resulted in acceptable mandibular range of motion compared with a 2.0 mm non locking plate.