Imaging quality evaluation of low tube voltage coronary CT angiography using low concentration contrast medium
Chengzhong Zhang1, Yuejun Yu2, Zaixian Zhang1
1Department of Radiology, Shanghai General Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, People's Republic of China.
Purpose:
To compare the image quality of prospectively ECG-gated low voltage coronary computed tomography angiography (CTA) with an administration of low concentration contrast medium.
Method And Materials:
A total of 101 patients, each with a heart rate below 65 beats per minute (BPM), underwent a prospectively ECG-gated axial scan in CT coronary angiography on a 64-slice CT scanner. All patients were allocated in three groups (group A: n=31, 80 kVp, 300 mgI/ml; group B: n=34, 100 kVp, 300 mgI/ml; group C: n=36, 120 kVp, 370 mgI/ml). The CT attenuation values of aortic root (AR), left main coronary artery (LMA), right main coronary artery (RMA) and chest subcutaneous fat tissue were measured. The contrast-to-noise ratio (CNR) of AR, LMA and RMA were calculated according to the formulas below. The values of computed tomography dose index (CTDI) and dose-length product (DLP) were recorded. Image quality was assessed on a 5-point scale. The results were compared using the one-way ANOVA and rank sum tests.
Results:
The values of CNR and SNR for vessels in group A and group B were not significantly different from group C (each p > 0.05). The effective radiation dose in group A (1.51 ± 0.70 mSv) and group B (2.59 ± 1.24 mSv) were both lower than group C (4.92 ± 2.82 mSv) (each p < 0.05). There was no significant difference among the image quality scores of group A (4.10 ± 0.41), group B (3.90 ± 0.48) and group C (4.04 ± 0.36) (each P > 0.05).
Conclusion:
Low tube voltage coronary CT angiography using low concentration contrast medium does not affect the imaging quality for assessing the coronary arteries compared with high voltage coronary CT angiography using high concentration contrast medium. Meanwhile low concentration contrast medium allowed 47-69% of radiation dose reduction.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...