Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Preliminary testing of GaN-based dosimeters for electron beam radiotherapy
1Protection and Safety Department, Atomic Energy Commission of Syria, P.B. 6091, Damascus, Syria Service de Cancérologie-Radiothérapie, CHU de Grenoble, CS 10217, Université Joseph Fourier Grenoble1, Grenoble Cedex 9 F-38043, France aismail@aec.org.sy.
Abstract:
The response of an implantable in vivo dosimetric system based on gallium nitride radioluminescence was investigated for electron beam radiotherapy using ELEKTA SLi and VARIAN Clinac 2100 CD Linear Accelerators. A bi-channel method has been implemented for fibre background rejection. The percentage depth dose (PDD) profiles were measured in polymethyl methacrylate for 6, 12 and 18 MeV electron beams. The PDD results were in excellent agreement with those measured with reference to ionisation chambers.