Related Experiment Video
Updated: Apr 15, 2026

Transcatheter Pulmonary Valve Replacement from Autologous Pericardium with a Self-Expandable Nitinol Stent in an Adult Sheep Model
Published on: June 8, 2022
Early outcomes of pulmonary valve replacement with the mitroflow bovine pericardial bioprosthesis
Sarah A Schubert1, John L Myers2, Allen R Kunselman3
1Department of Surgery, University of Virginia School of Medicine, Charlottesville, Virginia.
Background:
Bovine pericardial valves are often used for pulmonary valve replacement (PVR) in patients with previously repaired congenital heart disease. Attention has recently focused on the safety of the Mitroflow (Sorin Group USA, Arvada, CO) bovine pericardial valve after a national alert describing several cases of sudden valve failure in young patients. In response, we reviewed our experience using the Mitroflow bioprosthesis for PVR.
Methods:
Medical records were reviewed for all patients who underwent PVR using a Mitroflow valve at our center (2008-2013).
Results:
The cohort included 84 patients with a median age of 18.3 years (range, 0.8-62.1 years) and weight of 48.4 kg (range, 5.7-167.8 kg). Indications for surgical intervention included native outflow tract insufficiency (59 patients), valved conduit failure (20 patients), and isolated prosthetic valve failure (5 patients). Median length of stay was 3 days (range, 2-13 days). There were no hospital deaths. Median follow-up was 2.4 years (range, 0.2-5.6 years). Pulmonary valve insufficiency and peak gradient increased with time. At latest follow-up, freedom from insufficiency greater than or equal to a moderate degree was 83%, and freedom from a peak gradient greater than or equal to 50 mm Hg was 92%. Reintervention was required in 4 patients. One patient experienced endocarditis and underwent surgical PVR (0.6 years after initial intervention). Three patients underwent transcatheter treatment for valve stenosis including PVR (at 4.2 and 4.4 years in 2 patients) and balloon valvuloplasty (at 5.2 years in the third patient). Kaplan-Meier freedom from reintervention at 5 years was 81%.
Conclusions:
Early outcomes using the Mitroflow bovine pericardial valve for PVR in children and adults with repaired congenital heart disease appear acceptable and similar to reported outcomes for other tissue valve options. Valve failure from premature structural deterioration was not observed.