Objectives:
The purpose of the study is to compare real ear to coupler difference (RECD) curves based on physical and psychophysical measures. For the physically measured RECD, the RECD was measured with real ear and coupler measurements for the ear simulator and HA1- and HA2 2-cc couplers. The psychophysically measured RECDs were derived from audiogram measures.
Design:
RECDs were measured in 19 normally hearing subjects. The coupler measurement was done with the probe microphone and the coupler microphone itself. Psychophysically measured RECDs were derived for all subjects by measuring the audiogram in sound field and with an ER-3A insert phone.
Results:
Reference data were obtained for the three coupler types. It was possible to derive the RECD curve with psychophysical methods. There was no overall statistical difference between the physically and psychophysically measured RECD curves for the HA2 2-cc coupler and the ear simulator. The standard deviation was, however, much higher for the psychophysically derived RECD, indicating that physically measured RECDs are more precise than psychophysically derived RECDs.
Conclusions:
For the physical RECD measurements, the coupler microphone should be used for the coupler measurement. Physically measured RECDs were validated on group level by the reliable derivation of the RECD curve from audiogram measures.
More Related Videos
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
03:58Enhancing Electrode Location Assessment in Cochlear Implantation via Computed Tomography Image Fusion
Published on: January 17, 2025
Related Concept Videos
Perceiving Loudness, Pitch, and Location
Place theory, or place coding, suggests that different pitches are heard because various sound waves activate specific locations along the cochlea's basilar membrane. The brain determines the pitch of a sound by...
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Anatomy of the Ear
Auditory Perception