Resultant vertical prism in toric soft contact lenses
Anna Sulley1, Ryan Hawke2, Kathrine Osborn Lorenz3
1Johnson & Johnson Vision Care, Wokingham, UK.
Purpose:
Rotational stability of toric soft contact lenses (TSCLs) is achieved using a range of designs. Designs utilising prism or peripheral ballast may result in residual prism in the optic zone. This study quantifies the vertical prism in the central 6mm present in TSCLs with various stabilisation methods.
Method:
Vertical prism was computed using published refractive index and vertical thickness changes in the central optic zone on a full lens thickness map. Thickness maps were measured using scanning transmission microscopy. Designs tested were reusable, silicone hydrogel and hydrogel TSCLs: SofLens(®) Toric, PureVision(®)2 for Astigmatism, PureVision(®) Toric, Biofinity(®) Toric, Avaira(®) Toric, clariti(®) toric, AIR OPTIX(®) for ASTIGMATISM and ACUVUE OASYS(®) for ASTIGMATISM; with eight parameter combinations for each lens (-6.00DS to +3.00DS, -1.25DC, 90° and 180° axes).
Results:
All TSCL designs evaluated had vertical prism in the optic zone except one which had virtually none (0.01Δ). Mean prism ranged from 0.52Δ to 1.15Δ, with three designs having prism that varied with sphere power. Vertical prism in ACUVUE OASYS(®) for ASTIGMATISM was significantly lower than all other TSCLs tested.
Conclusions:
TSCL designs utilising prism-ballast and peri-ballast for stabilisation have vertical prism in the central optic zone. In monocular astigmats fitted with a TSCL or those wearing a mix of toric designs, vertical prism imbalance could create or exacerbate disturbances in binocular vision function. Practitioners should be aware of this potential effect when selecting which TSCL designs to prescribe, particularly for monocular astigmats with pre-existing binocular vision anomalies, and when managing complaints of asthenopia in monocular astigmats.
More Related Videos
Related Concept Videos
Focusing of Light in the Eye
Transformation of Plane Stress
Deformations in a Transverse Cross Section
As the material stretches, it expands or contracts in orthogonal directions to the load. This phenomenon varies...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Centroid for the Paraboloid of Revolution
The centroid for the paraboloid of revolution is the point where all the mass of the paraboloid is concentrated. This centroid is important for engineering applications, as it determines how forces are...