Single- versus multi-phase acquisition protocol for prospective-triggered sequential dual-source CT coronary
Wei Huang1, Yiming Xu2, Daoyan Lu2
1Department of Medical Imaging, Nanjing Jinling Hospital, Southern Medical University (Guangzhou), 305# Eastern Zhongshan Road, Nanjing 210002, China; Department of Medical Imaging, Huai'an First People's Hospital, Nanjing Medical University, 6# Western Beijing Road, Huai'an 223300, China.
Objective:
To investigate image quality and radiation dose of single- versus multi-phase acquisition protocol for prospective-triggered sequential dual-source computed tomography (CT) coronary angiography.
Materials And Methods:
A total of 140 patients were randomly assigned to single- or multi-phase group. Image quality and radiation dose were compared.
Results:
No significant difference was found in image quality between the two groups. Effective dose of single-phase group was 21.6% lower than that of multi-phase group (P<.001).
Conclusions:
Prospective-triggered sequential dual-source CT coronary angiography using single-phase protocol can reduce radiation dose without sacrifice of image quality in diastole compared with multi-phase protocol.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Acute Coronary Syndrome III: Diagnostic Studies
Cardiac Catheterization I: Pre-Procedure Overview
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT