VALIDATION OF AN INDEXED RADIOTHERAPY HEAD POSITIONING DEVICE FOR USE IN DOGS AND CATS
Katherine S Hansen1, Alain P Théon1, Sonja Dieterich2
1Department of Surgical and Radiological Sciences, VM: Surgery Radiology, University of California, Davis, CA, USA, 95616.
Abstract:
Setup variability affects the appropriate delivery of radiation and informs the setup margin required to treat radiation patients. Twenty-four veterinary patients with head and neck cancers were enrolled in this prospective, cross-sectional study to determine the accuracy of an indexed board immobilization device for positioning. Couch position values were defined at the first treatment based on setup films. At subsequent treatments, patients were moved to the previously defined couch location, orthogonal films were acquired, table position was modified, and displacement was recorded. The mean systematic displacement, random displacement, overall displacement, and mean displacement values of the three-dimensional (3D) vector were calculated. Three hundred thirty-two pairs of orthogonal setup films were analyzed for displacement in cranial-caudal, lateral, and dorsal-ventral directions. The mean systematic displacements were 0.5, 0.8, and 0.5 mm, respectively. The mean random displacements were 1.0, 1.1, and 0.7 mm, respectively. The overall displacements were 1.1, 1.4, and 0.9 mm, respectively. The mean 3D vector value was 1.6 mm with a standard deviation of 1.2 mm. Ninety-five percent of the vectors were <3.6 mm. These values were compared to data obtained with a previously used immobilization device. A t-test was used to compare the two devices. The 3D vector, random displacement in all directions, and overall displacement in the cranial-caudal and dorsal-ventral directions were significantly smaller than displacements with the previous device. The precision and accuracy of the indexed board device was superior to the historical head and neck device.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021