Air pressure-induced iridocornea contact in a patient with primary angle closure observed with a dynamic Scheimpflug
Rumi Kawashima1, Kenji Matsushita, Hisataka Fujimoto
1Department of Ophthalmology, Osaka University Graduate School of Medicine, Osaka, Japan.
Purpose:
To report air pressure-induced corneal deformation and iridocornea contact in eyes with primary angle closure (PAC) during intraocular pressure (IOP) measurement performed using a novel noncontact tonometer.
Methods:
A single case report.
Results:
We report a patient with bilateral angle closure. One eye had acute PAC and the other had PAC. The latter was evaluated by the movements of the cornea and iris during IOP measurement using a noncontact tonometer. During the examination, the corneal endothelium and the iris came into contact at the mid-peripheral pupillary area in the left eye with PAC during the corneal reaction to an air puff. In contrast, the corneal endothelium in the pupillary area did not come into contact with the iris.
Conclusions:
Although we observed only 1 case and there could be limitations in its interpretation, IOP measurements using a noncontact tonometer may create mechanical stress on the corneal endothelium in eyes with PAC with a very shallow anterior chamber.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
08:55Translaminar Autonomous System Model for the Modulation of Intraocular and Intracranial Pressure in Human Donor Posterior Segments
Published on: April 24, 2020
Related Concept Videos
Angle Closure Glaucoma: Treatment
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Focusing of Light in the Eye
Microbiome of the Eye
Ophthalmic Drug Delivery Systems