Radiation dose levels for conventional chest and abdominal X-ray procedures in elected hospitals in Sudan
E Babikir1, Hussein A Hasan2, A Abdelrazig3
1Radiological Sciences Department, College of Applied Medical Sciences, King Saud University, P.O. Box 10219, Riyadh 11433, Kingdom of Saudi Arabia Diagnostic Radiology Department, College of Medical Radiologic Sciences, Sudan University of Science and Technology, P.O. Box 1908, Khartoum, Sudan etom@ksu.edu.sa.
Abstract:
This study aimed to assess patient entrance surface air kerma (ESAK) during chest and abdominal X-ray procedures in screen film radiography (SFR) and computed radiography (CR) to establish dose reference levels. Patients' doses were measured in five hospitals for a total of 196 patients. ESAK was calculated from exposure parameters using DosCal software. The X-ray tube output (mGy mAs(-1)), accuracy of exposure factors, linearity and reproducibility were measured using an Unfors Xi dosimeter. The overall mean and range of ESAK during chest X-ray were 0.6 ± 0.3 (0.1-1.3) mGy, while for abdominal X-rays they were 4.0 ± 3.2 (1.3-9.2) mGy. Hospital with a CR system was found to use relatively higher doses. Dose values for abdominal X-ray procedures were comparable with previous studies. The dose for chest X-ray procedure was higher by a factor of 2-3 compared with the current international reference levels.