Atropine absorption after intramuscular administration with 2-pralidoxime chloride by two automatic injector devices
K E Friedl1, C J Hannan, P W Schadler
1Department of Clinical Investigation, Madigan Army Medical Center, Tacoma, WA 98431-5454.
Journal of Pharmaceutical Sciences
|September 1, 1989
Summary
The MARK I autoinjector significantly improves atropine absorption compared to other methods for treating anticholinesterase poisoning. This device enhances drug dispersal and separation, leading to faster and more effective treatment.
Area of Science:
- Pharmacology
- Toxicology
- Medical Devices
Background:
- Anticholinesterase poisoning requires rapid antidote administration.
- Manual intramuscular injection of atropine and pralidoxime chloride can lead to reduced drug absorption rates when co-administered.
- Automatic injectors offer potential for improved drug delivery and absorption.
Purpose of the Study:
- To compare the absorption rates of atropine and pralidoxime chloride when delivered via two different automatic injection devices.
- To evaluate the efficacy of the MARK I autoinjector in enhancing atropine absorption.
- To identify sensitive physiological markers of atropine action.
Main Methods:
- A crossover study involving 20 healthy young males (ages 20-30).
- Sequential intramuscular injection of citrated atropine and pralidoxime chloride (PAMCL) using either a multichambered autoinjector or the MARK I device (delivering drugs to separate sites).
- Assessment of atropine absorption via serum levels and physiological parameters (heart rate, salivary secretion, pupil diameter, near vision accommodation).
Main Results:
- The MARK I device resulted in significantly greater atropine absorption within the first 30 minutes post-injection.
- Salivary secretion changes were identified as a potentially sensitive and convenient marker for atropine's anticholinergic effects.
- The 6.9-mumol dose of atropine demonstrated near-maximal antisialogogue activity in the study population.
Conclusions:
- The MARK I autoinjector enhances atropine absorption, likely due to improved drug dispersal and separation from PAMCL.
- Salivary secretion is a practical and sensitive indicator of atropine's therapeutic effects.
- Automatic injectors represent an advancement in the delivery of antidotes for anticholinesterase poisoning.