Feasibility of ASL spinal bone marrow perfusion imaging with optimized inversion time
Dong Xing1, Yunfei Zha1, Liyong Yan1
1Department of Radiology, Renmin Hospital of Wuhan University, Wuhan, Hubei, China.
Journal of Magnetic Resonance Imaging : JMRI
|April 10, 2015
Summary
Flow-sensitive alternating inversion recovery (FAIR) magnetic resonance imaging is a reproducible and feasible method for measuring spinal bone marrow perfusion. This technique shows a strong correlation with dynamic contrast-enhanced MRI parameters.
Area of Science:
- Radiology
- Medical Imaging
- Physiology
Background:
- Spinal bone marrow (SBM) perfusion assessment is crucial for understanding various physiological and pathological conditions.
- Dynamic contrast-enhanced magnetic resonance imaging (DCE-MRI) is a common method for evaluating perfusion, but it involves contrast agents.
- Flow-sensitive alternating inversion recovery (FAIR) offers a potential alternative for SBM perfusion measurement without contrast agents.
Purpose of the Study:
- To evaluate the correlation between FAIR and DCE-MRI for SBM perfusion quantification.
- To determine the optimal inversion time (TI) for SBM FAIR perfusion imaging.
- To assess the reproducibility of the SBM FAIR perfusion technique.
Main Methods:
- FAIR perfusion imaging was performed on volunteers to optimize TI by measuring signal intensity changes (ΔM) and signal-to-noise ratio (SNR).
- Reproducibility was assessed through repeat SBM FAIR experiments in a subset of volunteers.
- Correlation analysis was conducted between FAIR-derived blood flow (BFASL) and DCE-MRI parameters (Ktrans, Kep, Ve) in subjects.
Main Results:
- An optimal TI of approximately 1200 msec was identified, maximizing ΔM and SNR for SBM FAIR perfusion.
- The SBM FAIR perfusion protocol demonstrated good reproducibility with a low within-subject coefficient of variation (3.28%).
- BFASL showed significant correlations with Ktrans (P < 0.001) and Kep (P = 0.004), but not with Ve (P = 0.082).
Conclusions:
- 1200 msec is the optimal TI for high-quality SBM ASL perfusion imaging using FAIR.
- The SBM FAIR perfusion protocol is reproducible and reliable for blood flow measurement.
- FAIR is a feasible technique for SBM blood flow assessment, showing strong correlation with DCE-MRI parameters.