Related Experiment Video
Updated: Apr 15, 2026

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
Versatile robotic probe calibration for position tracking in ultrasound imaging
Lars Eirik Bø1, Erlend Fagertun Hofstad, Frank Lindseth
1SINTEF Technology and Society, PO Box 4760 Sluppen, 7465 Trondheim, Norway. Norwegian University of Science and Technology, Faculty of Medicine, PO Box 8905, 7491 Trondheim, Norway. The Central Norway Regional Health Authority, PO Box 464, 7501 Stjørdal, Norway.
Abstract:
Within the field of ultrasound-guided procedures, there are a number of methods for ultrasound probe calibration. While these methods are usually developed for a specific probe, they are in principle easily adapted to other probes. In practice, however, the adaptation often proves tedious and this is impractical in a research setting, where new probes are tested regularly. Therefore, we developed a method which can be applied to a large variety of probes without adaptation. The method used a robot arm to move a plastic sphere submerged in water through the ultrasound image plane, providing a slow and precise movement. The sphere was then segmented from the recorded ultrasound images using a MATLAB programme and the calibration matrix was computed based on this segmentation in combination with tracking information. The method was tested on three very different probes demonstrating both great versatility and high accuracy.