Compatibility of medications during multi-infusion therapy: A controlled in vitro study on a multilumen infusion
Maxime Perez1, Bertrand Décaudin1, Aurélie Foinard2
1Department of Biopharmacy, Galenic and Hospital Pharmacy, Lille 2 University, UDSL, EA GRITA, UFR Pharmacie, 59037 Lille, France; Department of Pharmacy, CHRU Lille, 59037 Lille, France.
Objective:
Drug incompatibilities can jeopardize the safety and effectiveness of intravenous drug therapies, especially in the field of anaesthesia and intensive care. Patients receive many drugs simultaneously through limited venous accesses. This study was designed to confirm the impact of a multilumen infusion device on the occurrence of known physical drug incompatibilities.
Study Design:
In vitro laboratory work.
Methods:
Two infusion devices were studied: a standard single-lumen set and a multilumen infusion access device (Edelvaiss Multiline-8, Doran International). Up to six drugs were infused simultaneously: three acidic solutions of midazolam, amiodarone and dobutamine, and three alkaline solutions of furosemide, pantoprazole and amoxicillin/clavulanate. Saline, Ringer' solution and 5% dextrose were used as hydration vehicles with an infusion rate initially set at 100 mL/h and with stepwise decreases of 10 mL/h until precipitation. Two methods were used to highlight physical drug compatibility according to the European Pharmacopoeia: visual inspection of the extension set and an obscured-light sub-visible particle count test of infusions. The lowest infusion rate value for vehicle infusion to satisfy the two tests in all trials is reported for each infusion device.
Results:
The standard set did not satisfy the test in 82% of the assessed drug combinations. The Edelvaiss Multiline-8 was able to prevent the occurrence of drug incompatibilities in 49% of the drug combinations tested. This device is therefore advantageous, especially when simultaneously infusing two or four incompatible drugs.
Conclusions:
Infusion device characteristics have an impact on physical drug incompatibilities. Our results confirm that the Edelvaiss Multiline-8 device prevents physical drug incompatibilities under specified conditions.
Related Concept Videos
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Two-Compartment Open Model: IV Infusion
The model illustrates the decrease in plasma drug concentration from the central compartment with a specific equation. It shows that under steady-state conditions, the drug's input rate...
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
IV Infusion to Oral Dosing: Conversion Methods