Measurement and analysis of patient attenuation correction factor during radioiodine therapy
K Soliman1, A Alenezi2, M Bakkari2
1Department of Medical Physics, Prince Sultan Military Medical City, PO Box 7897, Riyadh 11159, Saudi Arabia soliman_k@hotmail.com.
Abstract:
The calculated dose rate from the radioiodine therapy patient should normally include a factor accounting for the attenuation and scatter of patient body tissues. The attenuation factor is currently neglected, and not applied in operational radiation protection. Realistic estimation of radiation dose rate levels from radioiodine therapy patients when properly performed will reduce operational cost and optimise institutional radiation protection practice. In this work, the existence of a patient body tissue attenuation factor is verified by comparing the dose rates measured from the radioiodine capsules immediately before administration with those measured from the patient immediately after administration. The correlation between the factors suspected to influence the patient body tissue attenuation and the measured dose rates from the patient normalised per unit activity is statistically analysed. The calculated attenuation correction factor based on authors' measurements was (0.55 ± 0.17). The measured dose rate per unit of radioactivity from the patient showed a negative correlation with their body mass index.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Radiological Investigation I: X-ray and CT