Related Experiment Video
Updated: Apr 15, 2026

A Contemporary Warming/Restraining Device for Efficient Tail Vein Injections in a Murine Fungal Sepsis Model
Published on: November 6, 2020
A resistive heating system for homeothermic maintenance in small animals
Veerle Kersemans1, Stuart Gilchrist1, Philip D Allen1
1Cancer Research UK and Medical Research Council Oxford Institute for Radiation Oncology, Department of Oncology, University of Oxford, Old Road Campus Research Building, Off Roosevelt Drive, OX3-7DQ, Oxford, United Kingdom.
A novel MR-compatible resistive heater using twisted pair wire effectively maintains animal body temperature during anesthesia and MRI scans. This heater avoids image artifacts, ensuring high-quality Magnetic Resonance Imaging (MRI) data.
Area of Science:
- Biomedical Engineering
- Medical Imaging
- Animal Research
Background:
- Maintaining stable body temperature in anesthetized animals is crucial for accurate physiological monitoring and Magnetic Resonance Imaging (MRI).
- Conventional heating methods can interfere with MRI data acquisition, causing artifacts and compromising image quality.
Purpose of the Study:
- To develop and evaluate an MR-compatible resistive heater for precise temperature maintenance in anesthetized animals.
- To assess the heater's impact on MRI image and spectral quality.
Main Methods:
- A resistive electrical heater was constructed using tightly-wound twisted pair wire and connected to a homeothermic controller.
- MRI scans (fat-suppressed images and localized spectra) were performed with the twisted pair heater and a single-strand wire heater.
- Data were acquired with and without current to assess MR insensitivity to heater-induced distortions.
Main Results:
- The twisted pair wire heater produced no discernible interference with MRI images or spectra.
- In contrast, a single-strand wire heater caused significant field shifts and lineshape distortions.
- Rectal temperature was effectively maintained, with lost temperature recovered within 10-20 minutes post-anesthesia induction.
Conclusions:
- The developed twisted pair wire heater is compatible with MRI environments.
- It effectively maintains animal body temperature without compromising MR image or spectral quality.
- This technology enables reliable physiological monitoring during MR studies.
Related Concept Videos
Decreased Body Temperature
Thermoregulation
Homeostatic Imbalances in Body Temperature
Body Temperature
Mechanism of heat transfer