Evaluation of 4 Commercial Viewing Devices for Radiographic Perceptibility and Working Length Determination
Trent Lally1, James R Geist2, Qingzhao Yu3
1Department of Endodontics, Louisiana State University Health Science Center, School of Dentistry, New Orleans, Louisiana.
Journal of Endodontics
|April 13, 2015
Summary
Desktop monitors and tablets performed similarly for detecting contrast and measuring working length, outperforming a laptop screen. This finding is crucial for accurate dental imaging interpretation across devices.
Area of Science:
- Radiology
- Dental Imaging
- Medical Device Evaluation
Background:
- Accurate interpretation of dental radiographs is crucial for diagnosis and treatment planning.
- Various display devices are used for viewing dental images, potentially impacting diagnostic quality.
Purpose of the Study:
- To compare the diagnostic performance of desktop monitors, laptop monitors, and tablets for contrast detection and working length measurements in dental imaging.
- To evaluate if differences exist between these display devices.
Main Methods:
- Radiographs of aluminum blocks and extracted teeth were created to simulate varying contrast and working length scenarios.
- Six observers assessed image quality and measurements on a desktop, a laptop, and two tablets.
- Statistical analyses, including ANOVA and kappa tests, were used to compare device performance and reliability.
Main Results:
- Desktop monitors and tablets showed comparable performance in contrast detection, significantly outperforming the laptop monitor.
- No significant differences were found among the four devices for working length measurement accuracy.
- Intraobserver and interobserver reliability were assessed.
Conclusions:
- While all devices were adequate for working length assessment, the laptop monitor exhibited inferior contrast perception.
- Display device choice can influence the ability to detect subtle contrast differences in dental images.
Related Concept Videos
X-ray Imaging
11.1K
German physicist Wilhelm Röntgen (1845–1923) was experimenting with electrical current when he discovered that a mysterious and invisible "ray" would pass through his flesh but leave an outline of his bones on a screen coated with a metal compound. In 1895, Röntgen made the first durable record of the internal parts of a living human: an "X-ray" image (as it came to be called) of his wife’s hand. Scientists worldwide quickly began their own experiments with...
11.1K
Determination of Crystal Structures
122
In the late 1800s, the revelation that light extended beyond visible wavelengths led to the discovery of X-rays by Wilhelm Roentgen. Recognized as high-energy electromagnetic radiation with short wavelengths, X-rays prompted exploration into their interaction with crystals. Max von Laue proposed in 1912 that the periodic arrangement of atoms, ions, or molecules in crystals would cause them to diffract X-rays, a hypothesis confirmed through experiments with copper sulfate and zinc sulfide...