Related Experiment Video
Updated: Apr 15, 2026

A Cost-effective and Reliable Method to Predict Mechanical Stress in Single-use and Standard Pumps
Published on: August 5, 2015
[Numerical assessment of impeller features of centrifugal blood pump based on fast hemolysis approximation model]
Abstract:
The impeller profile, which is one of the most important factors, determines the creation of shear stress which leads to blood hemolysis in the internal flow of centrifugal blood pump. The investigation of the internal flow field in centrifugal blood pump and the estimation of the hemolysis within different impeller profiles will provide information to improve the performance of centrifugal blood pump. The SST kappa-omega with low Reynolds correction was used in our laboratory to study the internal flow fields for four kinds of impellers of centrifugal blood pump. The flow fields included distributions of pressure field, velocity field and shear stress field. In addition, a fast numerical hemolysis approximation was adopted to calculate the normalized index of hemolysis (NIH). The results indicated that the pressure field distribution in all kinds of blood pump were reasonable, but for the log spiral impeller pump, the vortex and backflow were much lower than those of the other pumps, and the high shear stress zone was just about 0.004%, and the NIH was 0.0089.
More Related Videos
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
07:30In Vitro Model of Physiological and Pathological Blood Flow with Application to Investigations of Vascular Cell Remodeling
Published on: November 3, 2015