Comparison of Matrix with Humphrey Field Analyzer II with SITA
Marie-Josée Fredette1, Anik Giguère, Douglas R Anderson
1*MD †PhD ‡MD, MPH §OD, FAAO Department of Ophthalmology, Centre Universitaire d'Ophtalmologie, Université Laval, Centre Hospitalier Universitaire de Québec, Québec, le Québec, Canada (M-JF); CEVQ, SP-POS, Centre de Recherche du CHU de Québec, Québec, le Québec, Canada (M-JF, AG); Bascom Palmer Eye Institute, Department of Ophthalmology, Miller School of Medicine, University of Miami, Miami, Florida (M-JF, DRA, JMc); Office of Education and Continuing Professional Development, Department of Family and Emergency Medicine, Université Laval, Québec, le Québec, Canada (AG); and Department of Ophthalmology, University of North Carolina, Chapel Hill, North Carolina (DLB).
The Matrix perimeter and Humphrey Field Analyzer II (HFA) show different decibel values for contrast sensitivity. Matrix testing may benefit from duplicate measurements, especially for glaucoma patients.
Area of Science:
- Ophthalmology
- Visual Psychophysics
- Medical Device Technology
Background:
- Glaucoma diagnosis and monitoring rely on accurate visual field testing.
- Contrast sensitivity measurement is crucial for detecting early glaucomatous changes.
- Comparing new perimetry technologies with established standards is essential for clinical adoption.
Purpose of the Study:
- To compare the performance of the Matrix perimeter with the Humphrey Field Analyzer II (HFA).
- To evaluate contrast sensitivity measurements across different disease severities.
- To assess the repeatability and agreement between the two perimetry devices.
Main Methods:
- Fifty-three subjects (50 glaucoma, 3 normal) underwent visual field testing on both perimeters over five sessions.
- Intraclass correlation coefficients (ICC) for mean deviation (MD) and pattern standard deviation (PSD) were calculated.
- Repeatability of sensitivity estimates and agreement between instruments were analyzed using principal curve analysis.
Main Results:
- Both perimeters demonstrated high ICC (>0.90) for MD and PSD.
- A 1 dB difference on Matrix corresponded to a 2 dB difference on HFA for sensitivities above 20 dB.
- Matrix showed less dependence of measurement variability on disease severity compared to HFA, though high variability occurred at specific dB levels.
Conclusions:
- Decibel values from Matrix and HFA are not directly interchangeable.
- Matrix may offer more stable results across varying disease severities than HFA.
- Duplicate testing is recommended for Matrix, particularly for baseline and follow-up assessments in glaucoma management.
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Mass Analyzers: Common Types
Matrix-Assisted Laser Desorption Ionization (MALDI)
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences