In vitro study on work of breathing during non-invasive ventilation using a new variable flow generator
Rutger C Flink1, Anton H van Kaam2, Frans H de Jongh2
1Department of Neonatology, Emma Children's Hospital, Academic Medical Center, Amsterdam, The Netherlands Med-E Link, Amsterdam, The Netherlands.
A new Infant Flow SiPAP generator significantly reduces work of breathing (WOB) and improves pressure delivery in preterm infants. This innovation enhances nasal biphasic positive airway pressure (BPAP) support, potentially lowering respiratory failure risk.
Area of Science:
- Neonatal respiratory support
- Medical device engineering
- Pulmonary physiology
Background:
- Nasal synchronized biphasic positive airway pressure (BPAP) via the Infant Flow SiPAP system is increasingly used for preterm infants to reduce work of breathing (WOB) and respiratory failure risk.
- The existing generator's high resistance limits pressure amplitude (PA) and pressure build-up (PB), potentially compromising therapeutic effectiveness.
Purpose of the Study:
- To investigate the impact of a novel Infant Flow generator with enhanced fluid mechanics on WOB, PA, and PB during BPAP.
- To compare the performance of the new generator against the existing one in simulated clinical scenarios.
Main Methods:
- An in vitro low compliance lung model was utilized to measure WOB, PA, and PB.
- BPAP was delivered using both the old and new Infant Flow generators.
- Variations in airway resistance (tube sizes) and nasal interface sizes were employed, alongside different ventilator settings, to simulate diverse clinical conditions.
Main Results:
- The new generator demonstrated a significant reduction in WOB, ranging from 10% to 70% compared to the old generator.
- The maximum PA was substantially higher with the new generator (6-7 cmH2O) versus the old (3-4 cmH2O).
- The new generator achieved 33-40% of peak pressure within the first 100 ms of inspiration, compared to 11-20% for the old generator.
Conclusions:
- The new Infant Flow SiPAP generator significantly reduces WOB and enhances PA and PB during BPAP.
- These findings suggest improved respiratory support for preterm infants using the modified device.
- Further clinical validation in preterm infants under varied conditions is warranted to confirm these in vitro results.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...