Comparison of efficacy of three devices of manual positive pressure ventilation: a mannequin-based study
Somashekhar M Nimbalkar1,2, Suman Rao P N3, Saudamini V Nesargi4
1Department of Pediatrics, Pramukhswami Medical College, Karamsad, Anand, Gujarat, 388325, India. somu_somu@yahoo.com.
Background:
We compared the efficacy of and consistency in manual ventilation by trained healthcare professionals using three devices: self-inflating bag, flow-inflating bag, and T-piece resuscitator.
Methods:
Prospective analytical study at a level III Neonatal unit of a tertiary care hospital. Forty participants (consultants, postgraduates, interns, and neonatal nurses - 10 each) manually ventilated a mannequin with the above three devices for three minutes each. This procedure was video recorded. The pressure delivered during the three minutes and the breath rates for the first minute, second minute, and third minute were analyzed. Descriptive statistics were used to describe the study population and group statistics were used for various parameters of interest. Factorial analysis of variance was conducted to determine the main effects of device and specialty of users.
Results:
The mean (SD) peak inspiratory pressure of T-piece resuscitator was 16.5 (1.2), self-inflating bag (SIB) was 20.7 (4.4), and flow-inflating bag (AB) was 21.2 (5.0). The mean (SD) positive end expiratory pressure of T-piece resuscitator was 4.7 (0.9) cm of H2O and AB was 1.8 (1.7) cm of H2O. The maximum pressure delivered by T-piece resuscitator was 17.5, AB was 26.2, and SIB was 25.2 cm of H2O. Clinically appropriate breath rates were delivered using all of the devices. More effective breath rates were delivered using T-piece. There was no significant difference among the professional groups.
Conclusions:
The T-piece resuscitator provides the most consistent pressures and is most effective. Level of training has no influence on pressures delivered during manual ventilation.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: