Scattering correction algorithm in the PET sinogram using the factorial design of experimental method: A phantom
Huei-Yung Chen1, Nan-Han Lu2, Yung-Hui Huang3
1Department of Nuclear Medicine, E-DA Hospital, I-Shou University, Kaohsiung City, Taiwan.
Abstract:
Positron emission tomography (PET) had been utilized to image gene therapy, estimate tumor growth, detect neural function of the brain, and diagnose disease. However, sinogram noise always results inaccurate PET images. The factorial design of experiment (DOE), a statistical method, was applied to investigate, correct and estimate the fraction of scattering of 2D sinogram in PET. The DOE was included as factors of angle views and scatter media with two levels designed. The PET sinogram after scattering correction was then reconstructed by filtered back projection (FBP). Both Ge-68 uniform phantom and Jaszczak anthropomorphic torso phantom were applied to exam the performance of presented scattering correction algorithm. The signal-to-noise ratio (SNR), standard deviation (STD) of background, and full width at half maximum (FWHM), and uniformity test were applied to validate the performance of presented method. The proposed method provides a narrower FWHM, smaller STD of the background, higher SNR and better uniformity than those of original protocols. This method should be tested for accuracy and feasibility with three-dimensional phantoms or real animal studies and consideration effects of cross-talk between slices in future work.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022