1Department of Cardiology, Catharina Hospital, Michelangelolaan 2, 5623 EJ, Eindhoven, The Netherlands, f.bracke@me.com.
A restrained approach to replacing defibrillator leads with advisories is appropriate. Replacing leads during scheduled generator procedures may prevent future interventions, and lead extraction is not supported due to risks.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Safety
Background:
- Defibrillator lead advisories generate significant patient and physician concern, impacting management decisions.
- Previous advisories, such as for Medtronic Sprint Fidelis and St. Jude Riata leads, highlight potential complications.
- A need exists for an evidence-based approach to managing these lead advisories.
Purpose of the Study:
- To review the literature and provide an evidence-based approach to defibrillator lead management following advisories.
- To evaluate the risks and benefits of lead replacement versus abandonment.
- To inform clinical decision-making regarding compromised implantable cardioverter-defibrillator (ICD) leads.
Main Methods:
- Literature review focusing on complications associated with specific advisory leads (Medtronic Sprint Fidelis, St. Jude Riata).
- Analysis of consequences related to interventions, including lead replacement and extraction.
- Assessment of the risks associated with lead extraction versus abandonment.
Main Results:
- A restrained approach to premature replacement of advisory leads is recommended.
- Replacing leads during scheduled generator replacement may be opportune if electrical failure rates are elevated.
- No evidence supports the extraction of non-functional advisory leads; extraction risks often outweigh abandonment.
Conclusions:
- Clinical management of defibrillator leads with advisories should prioritize a conservative strategy.
- Scheduled generator replacement offers a potential window for proactive lead management in select cases.
- Abandoning non-functional leads is generally safer than attempting extraction due to procedural risks.