Related Experiment Video
Updated: Apr 14, 2026

Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
Intraocular lens power selection and positioning with and without intraoperative aberrometry
Purpose:
To determine the value of intraoperative aberrometry in cases of toric intraocular lens (IOL) implantation and positioning.
Methods:
In this non-randomized retrospective comparative trial, two groups of eyes underwent cataract extraction with toric IOL implantation: the aberrometry group (n = 37 eyes), where toric IOL power and alignment were determined before surgery with automated keratometry, standard optical biometry, and an online calculator and then refined using intraoperative aberrometry, and the toric calculator group (n = 27 eyes), where IOL selection was performed in a similar manner but without intraoperative aberrometry. The primary outcome measure was mean postoperative residual refractive astigmatism (RRA).
Results:
Mean RRA measured at follow-up after surgery was 0.46 ± 0.42 and 0.68 ± 0.34 diopters (D) in the aberrometry and toric calculator groups, respectively (P = .0153). A 75% and 57% reduction in cylinder was noted between preoperative keratometric astigmatism and postoperative RRA in the aberrometry and toric calculator groups, respectively (P = .0027). RRA of 0.25 D or less, 0.50 D or less, 0.75 D or less, and 1.00 D or less was seen 38%, 78%, 86%, and 95% of the time, respectively, in the aberrometry group and 22%, 33%, 74%, and 89% of the time, respectively, in the toric calculator group. These data show that the chance of a patient being in a lower postoperative RRA range increased when intraoperative aberrometry was used (P = .0130).
Conclusions:
Patients undergoing cataract extraction with toric IOL placement aided by intraoperative aberrometry were 2.4 times more likely to have less than 0.50 D of RRA compared to standard methods.
More Related Videos
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Focusing of Light in the Eye
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Glaucoma: Overview