Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
MPnRAGE: A technique to simultaneously acquire hundreds of differently contrasted MPRAGE images with applications to
Steven Kecskemeti1, Alexey Samsonov2, Samuel A Hurley3
1Waisman Center and Department of Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Purpose:
To introduce a new technique called MPnRAGE, which produces hundreds of images with different T1 contrasts and a B1 corrected T1 map.
Theory And Methods:
An interleaved three-dimensional radial k-space trajectory with a sliding window reconstruction is used in conjunction with magnetization preparation pulses. This work modifies the SNAPSHOT-FLASH T1 fitting equations for radial imaging with view-sharing and develops a new rapid B1 correction procedure. MPnRAGE is demonstrated in phantoms and volunteers, including two volunteers with eight scans each and eight volunteers with two scans each. T1 values from MPnRAGE were compared with those from fast spin echo inversion recovery (FSE-IR) in phantoms and a healthy human brain at 3 Tesla (T).
Results:
The T1 fit for human white and gray matter was T1MPnRAGE = 1.00 · T1FSE-IR + 24 ms, r(2) = 0.990. Voxel-wise coefficient of variation in T1 measurements across eight time points was between 0.02 and 0.08. Region of interest-based T1 values were reproducible to within 2% and agree well with literature values.
Conclusion:
In the same amount of time as a traditional MPRAGE exam (7.5 min), MPnRAGE was shown to produce hundreds of images with alternate T1 contrasts as well as an accurate and reproducible T1 map that is robust to B1 errors.