Head-to-head comparison between two screening systems for HBsAG, anti-HBc, anti-HCV and HIV combination immunoassays
M Schmidt1, A Jimenez2, A Mühlbacher3
1DRK-Blutspendedienst Baden-Württemberg, Hessen, Institut für Transfusionsmedizin und Immunhämatologie, Frankfurt, Germany.
Background:
Mandatory screening of blood donations for hepatitis B and hepatitis C viruses and human immunodeficiency viruses 1 and 2 requires assays with exceptional sensitivity and specificity. This study reports the results from a direct head-to-head comparison of the Elecsys HBsAG II, Elecsys Anti-HBc, Elecsys Anti-HCV II and Elecsys HIV combi PT immunoassays with the respective ABBOTT PRISM/Architect instrument immunoassays in a multicentre blood bank evaluation study.
Study Design And Methods:
Assay validation was performed in the blood screening laboratories of four blood bank centres in Austria, Germany, Spain and Thailand, where both first-time donor samples (approximately 6000 donors) and repeat donor samples (approximately 14,000 donors) were screened.
Results:
Of all screened donor samples, 93 (0.46%) were confirmed to be positive using assays from both manufacturers. The specificity of all immunoassays was >99.5% and was comparable between first-time and multiple-time donors. A direct comparison between the assays from Roche and ABBOTT according to Bland and Altman analysis demonstrated equivalent quality.
Conclusions:
These results suggest that the Elecsys immunoassays for HBV, HCV and HIV infection, with a comparative sensitivity of 100% and a specificity exceeding the common technical specification threshold of >99.5%, meet the stringent performance criteria stipulated for blood donor screening for these infectious agents. Significant differences in the specificity between first-time and repeat donors were not detectable.