Comparing Short-Duration Electro-Oculograms with and without Mydriasis in Healthy Subjects
C Türksever1, S Orgül1, M G Todorova1
1Department of Ophthalmology, University Hospital Basel, Switzerland (Chairman a. i.: Dr. sc. nat. Norbert Spirig).
Short electro-oculograms (EOG) provide reliable recordings without subject cooperation. Mydriasis does not impact EOG recording quality, making it a viable option for clinical assessments.
Area of Science:
- Ophthalmology
- Neuroscience
- Medical Technology
Background:
- Artifact-free electro-oculogram (EOG) recordings typically require subject cooperation.
- Evaluating short-duration EOG recording characteristics is crucial for efficient clinical practice.
- The impact of mydriasis on EOG recording quality in controls needs further investigation.
Purpose of the Study:
- To assess the recording characteristics of short-duration electro-oculograms (EOG).
- To compare the effect of mydriasis on EOG recordings in a control cohort.
- To determine if mydriasis influences the quality of EOG data.
Main Methods:
- Electro-oculogram recordings were conducted using a RETI-port gamma plus2 system with a light-emitting diode stimulus screen.
- Standardized light and dark phases with specific fixation durations and sweep counts were employed.
- Recordings were analyzed for Arden ratio, dark-trough, and light-peak values, with statistical comparisons made between conditions with and without mydriasis, and across age groups.
Main Results:
- Fifty-five control participants were categorized into three age groups (18-20, 20-40, 40-60 years).
- No significant differences were found in Arden ratio, dark-trough, or light-peak between recordings with and without mydriasis (p > 0.05).
- Age and tested eye did not significantly influence the recorded EOG parameters.
Conclusions:
- Short-duration electro-oculograms are effective for obtaining high-quality recordings.
- Mydriasis does not adversely affect the quality or reliability of electro-oculogram recordings.
- These findings support the use of short-duration EOG with mydriasis in clinical settings.