Sub-milliSievert (sub-mSv) CT colonography: a prospective comparison of image quality and polyp conspicuity at
Meghan G Lubner1, B Dustin Pooler, Douglas R Kitchin
1Departments of Radiology, University of Wisconsin School of Medicine and Public Health, E3/311 Clinical Sciences Center, 600 Highland Ave, Madison, WI, 53792-3252, USA, mlubner@uwhealth.org.
Objective:
To prospectively compare reduced-dose (RD) CT colonography (CTC) with standard-dose (SD) imaging using several reconstruction algorithms.
Methods:
Following SD supine CTC, 40 patients (mean age, 57.3 years; 17 M/23 F; mean BMI, 27.2) underwent an additional RD supine examination (targeted dose reduction, 70-90%). DLP, CTDI(vol), effective dose, and SSDE were compared. Several reconstruction algorithms were applied to RD series. SD-FBP served as reference standard. Objective image noise, subjective image quality and polyp conspicuity were assessed.
Results:
Mean CTDI(vol) and effective dose for RD series was 0.89 mGy (median 0.65) and 0.6 mSv (median 0.44), compared with 3.8 mGy (median 3.1) and 2.8 mSv (median 2.3) for SD series, respectively. Mean dose reduction was 78%. Mean image noise was significantly reduced on RD-PICCS (24.3 ± 19HU) and RD-MBIR (19 ± 18HU) compared with RD-FBP (90 ± 33), RD-ASIR (72 ± 27) and SD-FBP (47 ± 14 HU). 2D image quality score was higher with RD-PICCS, RD-MBIR, and SD-FBP (2.7 ± 0.4/2.8 ± 0.4/2.9 ± 0.6) compared with RD-FBP (1.5 ± 0.4) and RD-ASIR (1.8 ± 0.44). A similar trend was seen with 3D image quality scores. Polyp conspicuity scores were similar between SD-FBP/RD-PICCS/RD-MBIR (3.5 ± 0.6/3.2 ± 0.8/3.3 ± 0.6).
Conclusion:
Sub-milliSievert CTC performed with iterative reconstruction techniques demonstrate decreased image quality compared to SD, but improved image quality compared to RD images reconstructed with FBP.
Key Points:
• CT colonography dose can be substantially lowered using advanced iterative reconstruction techniques. • Iterative reconstruction techniques (MBIR/PICCS) reduce image noise and improve image quality. • The PICCS/MBIR-reconstructed, reduced-dose series shows decreased 2D/3D image quality compared to the standard-dose series. • Polyp conspicuity was similar on standard-dose images compared to reduced-dose images reconstructed with MBIR/PICCS.
More Related Videos
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Endoscopic Procedures II: Colonoscopy
Lower GI Series: Barium Enema
Procedure Details
The examination begins by inserting a lubricated rectal tube into the patient's rectum to administer a radiopaque barium solution. The barium flow is carefully...
Endoscopic Procedures III: Video Capsule Endoscopy
Imaging Studies III: Computed Tomography
Endoscopic Procedures IV: Sigmoidoscopy and Laproscopy
Sigmoidoscopy
Sigmoidoscopy is a diagnostic procedure that uses a flexible sigmoidoscope equipped with a light source and camera to examine the rectum and sigmoid colon. The procedure involves inserting the tube through the anus...