Pressures produced in vitro during intraligamentary anaesthesia
British Dental Journal
|November 25, 1989
Summary
The pressure syringe generated significantly higher pressures than the aspirating syringe during periodontal ligament injections. Male clinicians produced higher pressures with the aspirating syringe compared to female clinicians.
Area of Science:
- Dentistry
- Biomedical Engineering
Background:
- Periodontal ligament injections are crucial for dental anesthesia.
- Accurate pressure delivery is essential for effective and safe injections.
Purpose of the Study:
- To compare the pressures generated by an Astra type aspirating syringe and a Ligmaject pressure syringe during periodontal ligament injections.
- To investigate the influence of clinician sex and injection technique on pressure delivery.
Main Methods:
- An in vitro study utilizing a pressure transducer (0-6.9 MPa) connected to a microcomputer.
- Thirty-five clinicians (3:2 male:female ratio) performed periodontal ligament injections with both syringe types.
- Peak and time-averaged pressures were recorded and analyzed.
Main Results:
- The pressure syringe produced significantly higher pressures (P < 0.01) than the aspirating syringe.
- Male operators generated significantly higher pressures with the aspirating syringe (P < 0.1).
- No significant sex difference was observed with the pressure syringe (P > 0.1).
- A multiple high-pressure technique with the pressure syringe yielded significantly higher pressures (P < 0.01) than a steady pressure technique.
Conclusions:
- Pressure syringes are capable of generating higher pressures than aspirating syringes for periodontal ligament injections.
- Clinician sex may influence pressure delivery with aspirating syringes.
- Injection technique significantly impacts pressure output with pressure syringes.