Abstract:
A novel isocratic RP-HPLC method with UV photodiode array detection for the determination of seven organic acids in vinegar samples was developed and optimized. Samples were analyzed on a C18 (150 × 4.6 mm id, 5 μm) analytical column. Methanol-0.010 M sodium dihydrogen phosphate buffer adjusted with H3PO4 to pH 2.80 (2 + 98, v/v), was used as the mobile phase. The column oven temperature was optimized at 23.0°C, the flow rate was 0.80 mL/min, and the detection wavelength was 210 nm. Matrix-matched calibration curves were prepared for all analytes, and the correlation coefficients were greater than 0.99. LOD and LOQ ranged from 0.10 to 10.0 and 0.30 to 30.0 μg/mL, respectively. The results for interday and intraday precision and accuracy fell within the ranges specified. Vinegar samples were quantified by the standard addition method. The main advantages of this method are its selectivity, which is one of the main weaknesses of most methods when determining organic acids in complex matrixes, and its simplicity since little sample preparation is needed. Moreover, it is safe and inexpensive with low organic solvent usage.
More Related Videos
10:13Using Capillary Electrophoresis to Quantify Organic Acids from Plant Tissue: A Test Case Examining Coffea arabica Seeds
Published on: November 12, 2016
10:22Improved UPLC-UV Method for the Quantification of Vitamin C in Lettuce Varieties Lactuca sativa L. and Crop Wild Relatives Lactuca spp.
Published on: June 30, 2020
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Data Validation
Key parameters for method validation include:
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...
IR and UV–Vis Spectroscopy of Carboxylic Acids
However, the stretching absorptions for the C=O bond vary depending on the structure of carboxylic acids. The C=O bond of the free carboxylic acids shows a higher stretching frequency, 1760 cm−1, while H-bonded carboxylic acids (dimers) exhibit stretching absorptions at a lower frequency,...