Online Calibration of Glucose Sensors From the Measured Current by a Time-Varying Calibration Function and Bayesian
Goal:
Minimally invasive continuous glucose monitoring (CGM) sensors measure in the subcutis a current signal, which is converted into interstitial glucose (IG) concentration by a calibration process periodically updated using fingerstick blood glucose (BG) references. Though important in diabetes management, CGM sensors still suffer from accuracy problems. Here, we propose a new online calibration method improving accuracy of CGM glucose profiles with respect to manufacturer calibration.
Method:
The proposed method fits CGM current signal against the BG references collected twice a day for calibration purposes, by a time-varying calibration function whose parameters are identified in a Bayesian framework using a priori second-order statistical knowledge. The distortion introduced by BG-to-IG kinetics is compensated before parameter identification via nonparametric deconvolution.
Results:
The method was tested on a database where 108 CGM signals were collected for 7 days by the Dexcom G4 Platinum sensor. Results show the new method drives to a statistically significant accuracy improvement as measured by three commonly used metrics: mean absolute relative difference reduced from 12.73% to 11.47%; percentage of accurate glucose estimates increased from 82.00% to 89.19%; and percentage of values falling in the "A" zone of the Clark error grid increased from 82.22% to 88.86%.
Conclusion:
The new calibration method significantly improves CGM glucose profiles accuracy with respect to manufacturer calibration.
Significance:
The proposed algorithm provides a real-time improvement of CGM accuracy, which can be crucial in several CGM-based applications, including the artificial pancreas, thus providing a potential great impact in the diabetes technology research community.
Related Concept Videos
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Calibration Curves: Linear Least Squares
For data that follow a straight line, the standard method for fitting is the linear...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Glucose Homeostasis: Pancreatic Islets and Insulin Secretion
Insulin and C-peptide are...
Calibration Curves: Correlation Coefficient