Comparison of 4th-Generation HIV Antigen/Antibody Combination Assay With 3rd-Generation HIV Antibody Assays for the
Alagarraju Muthukumar1, Adnan Alatoom1, Susan Burns1
1Department of Pathology, University of Texas Southwestern Medical Center at Dallas.
Objective:
To assess the false-positive and false-negative rates of a 4th-generation human immunodeficiency virus (HIV) assay, the Abbott ARCHITECT, vs 2 HIV 3rd-generation assays, the Siemens Centaur and the Ortho-Clinical Diagnostics Vitros.
Methods:
We examined 123 patient specimens. In the first phase of the study, we compared 99 specimens that had a positive screening result via the 3rd-generation Vitros assay (10 positive, 82 negative, and 7 indeterminate via confirmatory immunofluorescent assay [IFA]/Western blot [WB] testing). In the second phase, we assessed 24 HIV-1 RNA-positive (positive result via the nuclear acid amplification test [NAAT] and negative/indeterminate results via the WB test) specimens harboring acute HIV infection.
Results:
The 4th-generation ARCHITECT assay yielded fewer false-positive results (n = 2) than the 3rd-generation Centaur (n = 9; P = .02) and Vitros (n = 82; P <.001) assays. One confirmed positive case had a false-negative result via the Centaur assay. When specimens from the 24 patients with acute HIV-1 infection were tested, the ARCHITECT assay yielded fewer false-negative results (n = 5) than the Centaur (n = 10) (P = .13) and the other 3rd-generation tests (n = 16) (P = .002).
Conclusions:
This study indicates that the 4th-generation ARCHITECT HIV assay yields fewer false-positive and false-negative results than the 3rd-generation HIV assays we tested.