Evaluation of test strips for the rapid identification of ethylenediaminetetraacetic acid (EDTA) specimens
Kelly L Scholes1, Sonia L La'ulu2, Ammon T Ence1
1ARUP Laboratories, Salt Lake City, Utah.
Background:
Identification of specimens that contain ethylenediaminetetraacetic acid (EDTA) is frequently necessary when investigating potentially mislabeled or improperly collected specimens.
Objective:
To evaluate the performance of rapid EDTA detection test strips in clinical specimens.
Methods:
We applied specimens to test strips designed to detect EDTA (QUANTOFIX EDTA) using a pipet (drop mode). Reactions were scored visually on a scale from red (no EDTA) to orange (low/indeterminate EDTA) to yellow (contains EDTA).
Results:
Test strips reliably identified specimens from EDTA-containing tube types. Although test strips did not detect strong reactivity in other specimens, tubes containing NaFl/K-oxalate produced an orange (low/indeterminate EDTA) reaction. Bismuth and citrate levels were higher in specimens after we dipped test strips into solution (dip mode).
Conclusions:
Test strips detected the presence of EDTA in concentrations found in EDTA-containing primary tubes. Test strips were less effective in evaluating low-level EDTA concentrations expected with intravenous line contamination or backflow. Indeterminate reactions required further investigation. Dip mode can produce analytical problems for assays that measure (or are interfered by) the contents of test strips.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Indirect and Alkalimetric Titration
Automated Microbial Diagnostics
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...